Joint replacements are the #1 expenditure of Medicare. The process of approving these medical devices is flawed according to the Institute of Medicine. It is time for patients' voices to be heard as stakeholders and for public support for increased medical device industry accountability and heightened protections for patients. Post-market registry. Product warranty. Patient/consumer stakeholder equity. Rescind industry pre-emptions/entitlements. All clinical trials must report all data.
Please share what you have learned!
Twitter: @JjrkCh
Showing posts with label St. Jude. Show all posts
Showing posts with label St. Jude. Show all posts

Sunday, September 9, 2012

Patient Last To Know About Implant Danger



September 7, 2012  The New York Times

Monsters attacked Avery de Groh when she was 4. That is how she remembers the day in 2007 when the defibrillator in her chest misfired, sending nine electric shocks through her body in less than 30 minutes.


Molly de Groh and her children Oliver, left, and Avery all have defibrillators implanted. When Avery’s misfired, repeatedly shocking her, doctors tried to fix it. Now they fear it could do it again.
Today, Avery is a chatty 9-year-old who just learned to roller-skate. She is old enough to know that she was not really attacked by monsters. The culprit was a broken wire from the defibrillator that keeps her heart beating normally. Like her mother and two brothers, she has an inherited condition that makes her prone to a fatal heart rhythm. After Avery’s episode, doctors removed the faulty wire, made by Medtronic, and replaced it with a new one made by St. Jude Medical.
Now it is possible that one is damaged, too. The wire, or lead, known as the Riata, was recalled in December after St. Jude warned doctors that internal cables were poking through the outer casing, causing unwanted shocks or failing to work when needed. Nearly 20 percent of the 128,000 people worldwide who have the Riata may be affected, according to the company.
Molly de Groh, Avery’s mother, said she worried that Avery’s new lead would also malfunction. “When I think about how scary it was for her,” she said, “I feel like, give that to me, and let her be fine.”
Heart device specialists have struggled for months to determine how best to treat patients with damaged leads. There is no easy fix: removing the wires can be dangerous, but so can leaving them in. In August, the Food and Drug Administration recommended that all patients with the Riata undergo imaging to see if their lead was failing. But the guidelines did little to settle the matter after some doctors questioned the wisdom of the advice.
Patients are caught in the middle, forced to grapple with life-or-death decisions for which there are no easy answers.
Mark Ulrich has decided not to wait. Several years ago, Mr. Ulrich’s defibrillator misfired in reaction to a medication he was taking. “I was pretty well barbecued,” said Mr. Ulrich, who is 68 and lives in Manhattan. “I would rather not be turned into a shish kebab.”
Dr. Jeffrey N. Rottman, a heart device specialist at Vanderbilt University, said most of his patients with damaged leads had elected to have them removed. “I think people who have a defibrillator already have a ‘just fix it’ type of approach,” he said.
That was Jacob Everidge’s attitude. Mr. Everidge, a 23-year-old from Athens, Ala., who has a condition that thickens the heart muscle, had his Riata removed by Dr. Rottman on Aug. 20. “I would much rather go ahead and get it out,” he said. “It wasn’t even a decision.”
For now, Ms. de Groh said, she will follow her doctor’s advice to wait and see if Avery’s lead begins to fail. But that raises other painful questions. Should the de Grohs cancel a family vacation in the Catskills, where they’ll be a half-hour from the nearest hospital? Can they send Avery to a sleepover? “As a mom of a fourth grader, I’ve got to send her to school,” said Ms. de Groh, a nurse in McHenry, Ill., a small town northwest of Chicago. But “for her not to be in my sight at all times is scary.”
Ms. de Groh and all three of her children were born with Long QT syndrome, which can cause their hearts to beat abnormally. She has a defibrillator made by Boston Scientific. Her 5-year-old son, Oliver, has the Durata, a newer lead made by St. Jude, whose safety has also been questioned recently. Her younger son, 3-month-old Monty, is too young for the operation, but he will eventually need a defibrillator, too. Because of her family history, Ms. de Groh has traveled twice to Washington to lobby lawmakers on device safety, but said she learned of the Riata recall only in August, after reading a newsletter written by a patient advocacy group.
She said she was angry at the F.D.A. and St. Jude for not contacting patients directly. “When something is implanted in a body, especially a child’s body, how can I find out about it through a newsletter?” she asked. Avery’s doctor, Marc Ovadia, said he chose not to tell Ms. De Groh about the recall because Avery’s lead is functioning “perfectly” and replacing it would require cracking open her chest. Telling patients about the recall, he said, could lead to unnecessary worry. “We want to make sure before we yell ‘fire’ in a crowded theater that this is fire,” he said. Still, he said, he regrets that Ms. de Groh found out about the recall the way she did. “These are tough, tough issues,” he said.
Mrs. de Groh’s frustration echoed that of consumer advocates who have criticized manufacturers and the F.D.A. for what they said was inadequate testing of medical devices before approval and a chaotic system for identifying problems once they are on the market.
In one example of the conflicting information about the devices, St. Jude reported last November that the problem with the Riata leads was affecting less than 1 percent of patients. But an internal report by an F.D.A. employee that month challenged that assessment, arguing that the company was underestimating the problem. The agency did not publicize the report, which was obtained through a Freedom of Information Act request and provided to The New York Times by a lawyer whose client is suing St. Jude.
The F.D.A. analysis proved to be correct: in July, a new St. Jude study found that the Riata showed signs of failing in 19 percent of patients.
The F.D.A. declined to comment on the report, other than to say that it was not unusual for the agency and the manufacturer to evaluate safety risks differently. Mitchell Shein, a manager in the F.D.A.’s division of cardiovascular devices, said deciding when to communicate with the public was often a tough call. “We’re very, very cautious to enter the public discourse on a medical issue unless we think we have something to add to that,” he said.
In addition to its recommendations that all patients with the Riata undergo imaging tests, the agency also ordered St. Jude to conduct additional studies on the Riata, which is no longer sold, and the Durata. Both were designed to be thinner than competing leads sold by Boston Scientific and Medtronic, making them easier to guide through blood vessels.
The Durata is made with a new coating that the company said seemed to have fixed the issue, but a study by a leading heart researcher has recently called that into question.
When a defibrillator is implanted, the lead is threaded through a blood vessel until it reaches the heart. In time, scar tissue can build up around it, making removal risky. The agency advised against removing the leads pre-emptively and said patients who had not experienced problems should undergo regular monitoring.
Some doctors have challenged the recommendation against routinely removing the leads, arguing that they can cause other problems when left in, like interfering with replacement leads. “There are potential risks associated with all options,” said Dr. Laurence Epstein, a heart device specialist at Brigham and Women’s Hospital in Boston.
Dr. Ovadia, a heart device specialist at Advocate Lutheran General Children’s Hospital outside of Chicago, said he did not know that the F.D.A. was recommending that all Riata patients undergo imaging tests until a reporter told him about the agency’s advice. And he said a St. Jude sales representative did not inform him of the new recommendations in a conversation in late August.
A St. Jude spokeswoman said the company notified doctors about the F.D.A. recommendations in a letter posted to its Web site, which it also sent to St. Jude field representatives. The letter explained the F.D.A.’s position on imaging, but also noted the issue was “complex and needs to take into account additional patient circumstances.”
In a separate statement, St. Jude said the company “continues to work closely with the F.D.A. and communicate important information with accuracy and integrity in a timely manner to inform patient care.”
Ms. de Groh said her thoughts had been turning to her baby, Monty. For now, he takes beta blockers to regulate his heartbeat. Common sense would tell her that he should receive a defibrillator when he is old enough. After all, medical devices fail in only a tiny percentage of cases.
“They’re supposed to save their life, but all it’s done in our family is cause problems,” Ms. de Groh said. “So you’re really conflicted as a parent on how to treat your kid if these devices are going to constantly be a source of fear and worry.
 

Saturday, May 12, 2012

Patient Harm: Riata ICD leads unreliable




Agonizing Choices for Heart Patients
Michael W. Minton has a device in his chest that could save his life by helping his heart beat properly. Or not.
A retired factory worker from Joelton, Tenn., Mr. Minton is one of 79,000 U.S. patients implanted with the troubled Riata defibrillator lead from St. Jude Medical Inc. STJ -0.64% The lead, a set of insulated wires, is threaded into the heart and connected to a defibrillator that zaps the heart into normal rhythm. But in a number of cases, the devices' wires are breaking through their insulation.
Doctors say the protruding wires could be a sign of dangerous electrical problems that could cause the lead to malfunction, shocking or even killing the patient.
Mr. Minton is one of those cases, though his device is still working properly, and now he's faced with a decision: Should he have the failure-prone heart device cut out, even though the procedure carries significant risk, or wait and hope that the Riata holds up.
After seeing an X-ray image of his lead that he said looked like an appliance wire stripped of its insulation, Mr. Minton, 58 years old, decided, "I want that thing out."
A small but growing number of patients are starting to agree. At a handful of institutions, including Vanderbilt University Medical Center, Brigham and Women's Hospital and the Minneapolis Heart Institute, doctors are starting to remove even some functioning Riata leads when the devices show signs of wear. So far, Vanderbilt doctors have used X-rays to identify 29 leads with significant insulation problems, and removed one-third of them, beginning with those that are also malfunctioning.
The full scope of the problem is unknown, though by some counts, involving small groups of patients, as many as one-third of Riata's leads have protruding wires. That lack of clear information creates "much uncertainty for physicians, patients and patients' families," wrote a group of U.S. senators in a May 7 letter seeking details of St. Jude's handling of the Riata situation. St. Jude said it is responding to the letter.
link

St. Jude's share price plunged 10% after public attention focused on the Riata leads a month ago. The stock fell 25 cents to $38.88 in 4 p.m. trading Friday on the New York Stock Exchange. Heart-rhythm devices, including defibrillators and leads, made up more than 50% of St. Jude's $5.6 billion in 2011 sales.
The procedure to remove leads, which can tear heart tissue or leave broken pieces of the wire inside veins, is controversial. Guidance from St. Jude, which stopped selling Riata-series leads in 2010, recommends against removing the devices before they malfunction, because the risk of complications is considered high. One study of a favored lead-removal technique, in 2010, found a 1.5% chance of major complications, which can include death.
St. Jude said in a statement that its guidelines were devised with help from an independent physician advisory panel. Physician opinions may differ, and clinical decisions in general should be made on a patient-by-patient basis, it said.
The company will release a 700-patient study in June that could offer doctors more information about the risks of leaving leads with insulation flaws in place. Each patient will be X-rayed to look for the protruding wires, a St. Jude spokeswoman said.
The Heart Rhythm Society, a medical group, hasn't advised doctors whether to remove the leads, and is awaiting further data. Its president, the Cleveland Clinic's Bruce L. Wilkoff said he has removed some leads that didn't have electrical problems.
Defibrillator leads are often bound to the walls of veins and the heart by scar tissue, an obstacle that worsens with time. To remove them, doctors slide sheaths over the wires and scrape the scar tissue away, freeing the lead. The tip of the wires can become anchored in the heart wall, and removing them can tear a hole.
At Vanderbilt, Mr. Minton's doctor, Christopher R. Ellis, said his group decided to recommend to patients the removal of all leads with frayed wires visible on X-rays, even if they work.
Dr. Ellis acknowledged that the Vanderbilt approach is aggressive, but he said the group's rate of complications appears lower than the risk of Riata problems. "I don't see anything good that can happen from leaving all these leads in," he said.
Riata was recalled in December after St. Jude's medical advisers noted a higher-than-normal rate of an insulation failure, called externalization. The leads are also prone to other types of failures, and in March a study linked the devices to 20 deaths caused by short circuits. Not all externalized leads also short circuit, but new data suggests it is a sign of the problem.
In research presented Thursday at a Heart Rhythm Society meeting in Boston, 110 patients at seven hospitals who were checked with X-rays showed a 25% rate of externalized leads. Nearly one-third of those leads also had electrical problems, said Raed Abdelhadi, a Minneapolis Heart Institue doctor who presented the finding. Vanderbilt contributed to the study.
If future studies confirm those problem rates, Dr. Abdelhadi said, "If I were a patient, I'd know what I'd want."
Doctors should await additional data, said Raymond H.M. Schaerf, a Burbank, Calif., surgeon who removes leads, but who has counseled his patients against extractions when leads are functioning properly.
"When doctors get over aggressive, patients suffer more from the extraction than the actual lead," Dr. Schaerf said.
In previous recalls, such as the 2007 recall of Medtronic Inc.'s MDT -1.02% Sprint Fidelis lead, more patients were harmed by overzealous doctors than faulty leads, he said. It is not yet clear how Riata will compare with Fidelis, he said.
Laurence Epstein, chief of cardiac arrhythmia at Brigham and Women's Hospital in Boston, said that because the true risks of Riata are unclear, he leaves the choice up to his patients. After seeing loose wires in their hearts in X-ray images, some choose the extraction, Dr. Epstein said. "People don't like that time-bomb thing," he said.
"If it is externalized, in my mind, the lead has failed," said Dr. Epstein, who like others called for more data to gauge the risk.By nature, Mr. Minton, Dr. Ellis's patient, said he was inclined to deal with his Riata problem upfront, even if the procedure brings its own risks. "I'm not the type to stand in the closet for a couple of hours because there is a tornado warning," he said.
Write to Christopher Weaver at christopher.weaver@wsj.com
A version of this article appeared May 12, 2012, on page B1 in some U.S. editions of The Wall Street Journal, with the headline: Agonizing Choices for Heart Patients.
 link